Abstract:Objective To evaluate the performance of ADVIA2120for the test of reticulocytes. Methods ADVIA2120was used for detection of high/normal/low reticulocyte samples and the reproducibility,the rate of carrying contamination and stability were observed.The results were compared with manual method. Results Reproducibility:except that the CV of high nucleic acid was13.82%~17.4%,the CV of the other indexes was less than8.19%;linearty:r=0.9979;stability:percentage CV of Ret at4℃in24hwas3.8%;the rate of carrying contamination was0.22%;compared with the manual methods:r=0.9962,P>0.05. Conclusion The srate of carrying contamination by using ADVIA2120blood-counter system in test of reticulocytes is low,stable and showing a nice linear correlation suitable for clinical use.
Key words:Reticulocytes;Automatic blood-counter system;Plasmodium
网织红细胞(Reticulocytes,Ret)传统的人工测定烦琐而费时,准确性受操作人员的经验、水平、细胞计数量、染色效果及涂片质量等因素影响大,重复性也差。全自动血细胞分析仪测定网织红细胞应用逐渐增多,因临床应用时间不长,其性能尚有待观察。ADVIA2120全自动血细胞分析仪是拜耳公司新推出的五分类机型,配备全自动网织红细胞分析功能,可提供14项相关参数,因部份指标临床意义尚无确论,故本文仅选择6项与细胞数量相关的指标进行评价。现进行重复性、线性、携带污染率、稳定性测试,并与手工百分率分类结果进行比较,报道如下。
1 资料与方法
1.1 资料
1.1.1 标本来源 住院患者,网织红细胞按仪器初步测量结果分为高、中、低三类。
1.1.2 设备 美国ADVIA2120全自动血细胞分析仪、日本OLYMPUSBX-60多功能光学显微镜、上海医疗设备厂HH-W21-600电热恒温水育箱、姜堰康健医疗器具有限公司KJRM-1摇匀器。
1.1.3 ADVIA2120全自动血细胞分析仪原装配套试剂及全血质控品(批号0605204);珠海贝索公司网织红细胞染液和瑞氏染液,按试剂使用说明操作。广州阳普公司EDTA-K2血常规抗凝管(批号0602012)。
1.2 方法 按相关评价原则,ADVIA2120全自动血细胞分析仪重复测定样本,评价仪器日内精密度、日间精密度、线性、携带污染率和稳定性,测定前摇匀器正、反向摇动3min。
1.2.1 日内精密度 按美国临床实验室标准化委员会(NC-CLS)制定的方案 [1] ,选择本院住院患者血样,网织红细胞按仪器初步测量结果分为高、中、低三类,各1份,每份重复测定10次,并计算平均值(x)和变异系数(CV)。
1.2.2 日间精密度 因临床血标本稳定性差,故以原厂全血质控品每日9:00、12:00、16:30测定三次,取均值,连续测定10d,并计算平均值(x)和变异系数(CV)。
2 结果
2.1 日内精密度测试 高值项目日内精密度较好,低值项目高核酸浓度网织红细胞比率(%)日内精密度偏大。见表1。
表1 ADVIA2120网织红细胞测定日内精密度(略)
2.2 日间精密度测试 高值项目日间精密度较好,低值项目高核酸浓度网织红细胞比率(%)日精密度偏大。见表2。
表2 ADVIA2120网织红细胞测定日间精密度(略)转贴于 酷文网-论文下载中心 http://www.coolwen.net